SHP1-V or SHP1-A SHP1-V or SHP1-A
Radiometer connection
up to
serial number 144999
wire from power supply ground
to Modbus® ground needed
Radiometer connection
starting from
serial numbers 145000
.
Please follow the instructions in this section carefully for the mechanical and electrical installation of the SHP1 pyrheliometer.
Do not turn on power to the instrument until instructed to do so.
Note Do not connect the instrument to a computer until instructed to do so.
Note Do not turn on power to the operating computer until instructed to do so.
2.1 Included with the product
Check the contents of the shipment for completeness (see below) and note whether any damage has occurred during transport. If
there is damage, a claim should be filed with the carrier immediately. In the case of damage and/or the contents are incomplete,
contact your local Kipp & Zonen representative or e-mail the Kipp & Zonen customer and product support department at:
support@kippzonen.com
Although a SHP1 pyheliometer is weather-proof and suitable for use in harsh environmental conditions, it has some delicate
mechanical parts. Please keep the original packaging for safe transport of the radiometer to the measurement site, or for use
when returning the radiometer for calibration.
The following items are included with a SHP1 pyrheliometer:
Smart pyrheliometer and rain shield
Cable, pre-wired with 8-pins connector or connector only for customer cable
Calibration certificate
Instruction sheet
2 Dessicant bags
CD with product documentation and software
2.2 Tools required
The tool required to mount a SHP1 on a SOLYS 2 or 2AP sun tracker is a 3 mm Allen key. Normally, the drying cartridge for the
SHP1 should be hand-tight, but a 16 mm or 5/8” open-ended wrench / spanner can be used to loosen it.
Check the condition of the desiccant in the SHP1 and replace before installation, if necessary; for example after a long storage
period.
The SHP1 instruction sheet plus the sun tracker manual contain all information to do the installation. When using the digital
output it might be convenient to set the Modbus® address prior to visiting the site, otherwise a computer and RS-485 / USB
converter may be required during installation.
2.3 Location and support
The following steps must be carefully taken for optimal performance of the instrument.
2.3.1 Location
Ideally, the site for the pyrheliometer plus sun tracker should be free from any obstructions to the hemispherical view from the
plane of the detector. If this is not possible, the site should be chosen in such a way that any obstruction over the azimuth range
between earliest sunrise and latest sunset should have an elevation not exceeding 5 ° (the apparent sun diameter is 0.5 °).
Further details for installation of the sun tracker can be found in the manual of the used tracker.
It is evident that the radiometer should be located in such a way that a shadow will not be cast upon it at any time (for example
by masts or ventilation ducts). Note that hot exhaust gas (> 100°C) will produce some radiation in the spectral range of the
radiometer and cause an oset in the measurements. This is important for an accurate measurement of the direct solar radiation.
The radiometer should be readily accessible for cleaning the front window and inspecting the desiccant.
2.3.2 Mounting
The mounting of the SHP1 pyrheliometer is related to the used sun tracker. Therefore we refer to the sun tracker manual for
further instructions on how to mount the SHP1 on the side mounting plate of the sun tracker.
2.3.3 Fitting the connector and cable
Locate the plug correctly in the radiometer socket, it only fits one way, and push it in. Screw the plug locking ring hand-tight.
Over-tightening may damage the waterproof seal. Secure the cable so that it cannot blow in the wind or cause a shadow on the
instrument.
Note The cable should be arranged with a curve below the instrument so that water drips o, rather than running along the
cable up to the connector.
2.4 Electrical connections
As standard SHP1 pyrheliometers are supplied with a waterproof connector pre-wired to 10 m of high quality yellow cable with 8
wires and a shield covered with a black sleeve. Longer cables are available as options. The colour code of the wires and the
connector pin numbers are shown below and on the instruction sheet.
Special attention is needed to prevent power or ground loops when connecting the SHP1 to multiple readout devices.
Connecting the RS-485 to a grounded circuit and the analogue output to a floating circuit can cause unacceptable
ground loops. This may cause dierential voltages outside the SHP1 specifications and will damage the unit. We
recommend using either the analogue or the digital output but not both. The maximum dierential between either
of the Modbus® RS-485 lines (yellow and grey) and the power ground / RS-485 common line (black) is 70 VDC.
First connect all wires before plugging into the radiometer
The shield of the cable is connected to the aluminium radiometer housing through the connector body. The shield
at the cable end may be connected to ground at the readout equipment. Lightning can induce high voltages in the
shield but these will be led o at the pyrheliometer and data logger.
Note Long cables may be used, but the cable resistance must be smaller than 0.1 % of the impedance of the readout
equipment for the analogue outputs and may aect the baud rate of the RS-485 digital connection.
2.5 Power connection
The minimum power supply voltage for a SHP1 pyrheliometer is 5 VDC. However, for optimal performance it is advised to use
12 VDC, especially when long cables are used. 5-volt power can only be used in combination with a short cable, maximum 10 m.
It is advised to protect the output of the power supply with a fast blowing fuse of maximum 250 mA rating.
.
Maximum power consumption and input current.
65 mW and 2 mA at the highest input voltage.
63 mW and 12.5 mA at the lowest input voltage.
The maximum inrush current is 200 mA.
The above mW values represent the dissipation within the SHP1-A. For the total power the energy in the load resistor has to
be added.
For supply voltages below 12 Volts or above 20 Volts it is advised to use a load resistor of less than 500 Ω to keep the power
consumption as low as possible.
2.6 Data connection
Connection to a Personal Computer by Universal Serial Bus (USB)
The connection depends on the use of a RS-485 to USB converter.
The converter must have galvanic isolation between the inputs and outputs to prevent possible damage to the SHP1
digital interface. This is particularly an issue with portable computers (laptops, etc.) in which the power supplies
can generate large voltage spikes.
A suitable converter is the model USOPTL4 from B & B Electronics. One end has the USB connector to the PC the other end has
a connector with screw terminals for the instrument wires. This RS-485 converter is powered from the USB interface, so no
additional power adaptor is necessary.
.
*Note Switches on the converter should be set for RS-485, 2-wire operation and Echo o.
.
*Note Switches on the converter should be set for RS-485, 2-wire operation and Echo o.
Connection to a RS-485 Network
The digital interface can be connected to a 2-wire RS-485 network as shown below.
The interface needs also an external power to provide the voltage for the electronics. If the interface is the last device on the
network then a terminator consisting of a 120Ω or 150 Ω resistor has to be connected between terminal A/A'/- and B/B'/+. Never
place the line termination on the derivation cable. It is required to install the pull up and pull down resistors as shown in the
previous figure. The value of these resistors has to be within 650Ω and 850Ω.
2.7 Analogue output connection
The SHP1-V (Volt version) has been factory set to an output of -200 to 2000 W/m². This applies only to the analogue output and
means that an output of 0 Volt corresponds to -200 W/m² (this will never be reached) and 1 Volt corresponds to 2000 W/m².
The digital output range can be modified with Modbus® commands. For the SHP1 the output range can be set to -200 to 4000 W/m²
for 0 to 1 Volt.
Typical power consumption SHP1-V
5 VDC max. 50 mW (approx. 10.0 mA)
12 VDC max. 55 mW (approx. 4.5 mA)
24 VDC max. 60 mW (approx. 2.5 mA)
The range has to start negative in order to show (small) negative readings also the analogue output itself cannot go negative. If
used in atmospheric conditions it is advised to keep the range as factory set.
The same applies for the SHP1-A (current version) that has been factory set to 0 to 1600 W/m² for 4 to 20 mA.
Here negative inputs will make the output go under 4 mA.
2.8 Calculations
2.8.1 Calculation 0 to 1 Volt version
The output is defined from 0 to 1 Volt representing -200 to 2000 W/m².
The irradiance value (
E
solar
) can be simply calculated as shown below in formula 1. The formula assumes the factory default setting
of the analogue output. For calculation of the solar irradiance (global or reflected) the following formula must be applied:
formula 1
E
solar
= Solar radiation [W/m²]
V = Output of radiometer [Volt]
2.8.2 Calculation 4 to 20 mA version
The output is defined from 4 to 20 mA representing 0 to 1600 W/m².
Negative outputs can cause the output to go slightly below 4 mA.
The irradiance value (
E
solar
) can be simply calculated as shown below in formula 2. The formula assumes the factory default setting
of the analogue output. For calculation of the solar irradiance (global or reflected) the following formula must be applied:
formula 2
E
solar
= Solar radiation [W/m²]
mA = Output of radiometer [mA]
2.8.3 Recommended cable types
Where cables need to be extended, or the customer prefers to provide their own cables, they should be suitable for outdoor used
and UV resistant.
Radiometer Connection
Wire Function Connect with
Red
Blue
HousingShield
Not connectedNone
Modbus® common / Ground
Analogue out V+/4-20 mA(+)
Analogue ground V
-
/4-20 mA(
-
)
Modbus® RS-485
Modbus® RS-485
Power 5 to 30 VDC
(12 V recommended)
Power ground / RS-485 Common
Ground *
White
Black
Yellow
Brown
Green
Grey
5
1
2
8
7
* Connect to ground if radiometer not grounded
The blue wire is not connect with radiometers
with serial number up to 144999
4
6
3
5 to 30 VDC
power supply
Modbus® RS-485
5 to 30 VDC
power supply
Modbus® RS-485
V / mA
0.0
0.2
0.4
0.6
0.8
1.0
V / mA
0.0
0.2
0.4
0.6
0.8
1.0
13