Order-No: S32100
S32200
S32300
PTB 110.3A
PTB 110.2A
PTB 110.1A
800 ... 1100 hPa
600 ... 1100 hPa
500 ... 1100 hPa
• Accuracy ±0.3 hPa @ +20 °C
• Delivery including IEC 17025 calibration certicate
Barometric Pressure Sensor Vaisala PTB110
The Vaisala BAROCAP® Barometer PTB110 is designed both for accurate
barometric pressure measurements at a room temperature and for gen-
eral environmental pressure monitoring over a wide temperature range.
Vaisala BAROCAP® technology
The PTB110 barometric pressure sensor uses the Vaisala BAROCAP® sen-
sor, a silicon capacitive absolute pressure sensor developed by Vaisala for
barometric pressure measurement applications. The sensor combines the
outstanding elasticity characteristics and mechanical stability of single-
crystal silicon with the proven capacitive detection principle.
Your advantages
• Several pressure ranges,
from 500 / 600 / 800 ... 1100hPa
• Accuracy ±0.3 hPa at +20 °C
• Long-term stability
• On / off control with external trigger
• Output voltage 0 ... 5 VDC
• Current consumption less than 4 mA
• Mountable on a (35 mm wide) DIN rail
• NIST traceable (certicate included)
PTB110
BAROMETRIC
PRESSURE
BAROCAP® is a registered trademark of Vaisala.
Specifications are subject to change without prior notice.
©Vaisala Oyj
Dimensions
Dimensions in mm [inches]
General
Supply voltage 10 ... 30 VDC
Suppply voltage control with TTL level trigger
Supply voltage sensitivity negligible
Current consumption less than 4 mA
in shutdown mode less than 1 µA
Output voltage 0 ... 2.5 VDC
0 ... 5 VDC
Output frequency 500 ... 1100 Hz
Resolution 0.1 hPa
Load resistance minimum 10 kohm
Load capacitance maximum 47 nF
Settling time 1 s to reach full accuracy after
power-up
Response time 500 ms to reach full accuracy
after a pressure step
Acceleration sensitivity negligible
Pressure connector M5 (10-32) internal thread
Pressure fitting barbed fitting for 1/8”
Minimum pressure limit 0 hPa abs
Maximum pressure limit 2000 hPa abs
Electrical connector removable connector for
5 wires (AWG 28 ... 16)
Terminals Pin 1: external triggering
Pin 2: signal ground
Pin 3: supply ground
Pin 4: supply voltage
Pin 5: signal output
Housing material, plastic cover ABS/PC blend
Housing classification IP32
Metal mounting plate Al
Weight 90 g
Electromagnetic compatibility Complies with EMC
standard EN 61326-1,
Generic Environment
Accuracy
Linearity* ±0.25 hPa
Hysteresis* ±0.03 hPa
Repeatability* ±0.03 hPa
Calibration uncertainty** ±0.15 hPa
Accuracy at +20 °C*** ±0.3 hPa
* Defined as ±2 standard deviation limits of end-point
non-linearity, hysteresis error or repeatability error.
** Defined as ±2 standard deviation limits of inaccuracy of the
working standard including traceability to NIST.
*** Defined as the root sum of the squares (RSS) of end-point
non-linearity, hysteresis error, repeatability error and
calibration uncertainty at room temperature.
Total accuracy at
+15 ... +25 °C ±0.3 hPa
0 ... +40 °C ±0.6 hPa
-20 ... +45 °C ±1.0 hPa
-40 ... +60 °C ±1.5 hPa
Long-term stability ±0.1 hPa/year
Ref. B210681EN-A 2007-04
Operating range (1 hPa=1mbar)
Pressure ranges 500 ... 1100 hPa
600 ... 1100 hPa
800 ... 1100 hPa
800 ... 1060 hPa
600 ... 1060 hPa
Temperature range -40 ... +60 °C
Humidity range non-condensing
Technical Data
13.40
0.53
68.40
90
3.54
4
0.16
76.20
3.00
3.50
0.14
Ø 7.20
0.28
15.82
0.62
97.3
3.83
28.1
1.11
Dimensional drawing
Accuracy and stability
The excellent long-term stability of the barometric pressure sensor minimizes or even removes the need for eld ad-
justment in many applications.
Applications
The PTB110 is suitable for a variety of applications, such as environmental pressure monitoring, data buoys, laser
interferometers, and in agriculture and hydrology. The compact PTB110 is especially ideal for data logger applications
as it has low power consumption.
page 1/3